BIOLOX®delta - 4th-generation ceramic-on-ceramic bearings, stability, and low complication rates: a study of 10+ years
The investigation specifically addresses concerns associated with the thinner liners required to accommodate larger femoral heads, namely liner fracture risk, late dislocation, noise, and long-term biological response.
From April 2010-February 2012, 246 patients (274 THAs) were enrolled. At final follow-up, 178 patients (201 hips; mean age 51 years; range 18.9-77.6) were assessed at a mean of 12.3 years (range 10-14.6), with CT scans performed in 47 hips (23%). Kaplan-Meier survival analysis used revision and reoperation as endpoints.
Findings to note |
|
Limitations, as noted by the authors: the study lacked a control cohort for comparison with other bearings, and the dropout rate was relatively high (26.7%, 73 of 274); only 23% of the cohort underwent CT at 10-year follow-up.
The authors conclude that large-head BIOLOX®delta CoC bearings can prevent early dislocation without incurring liner-fracture risk, even with the thinner liners that larger heads require. LDK-associated late instability and articular noise remain considerations in long-term follow-up.
📖 Lee DH, Park JW, Jung WH, Kim TY, Ha YC, Koo, KH, Lee, YK. Complications of ceramic-on-ceramic total hip arthroplasty with large heads: 10-year minimum follow-up. J Arthroplasty. 2026. doi: 10.1016/j.arth.2026.06.015
Please check for regulatory approval in your country.
This text reflects CeramTec’s summary of a peer-reviewed scientific publication and does not constitute clinical guidance, risk prediction for individual patients, or product-related recommendations. For product, safety, and risk information, always refer to the labeling of the legal manufacturer. This post was drafted with AI assistance and approved by CeramTec.